Vega VEGABAR 81 HART and accumulator pack - Operating I Bedienungsanleitung Seite 46

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 68
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 45
46
9 Diagnosis, asset management and service
VEGABAR 81 • HART and accumulator pack
45026-EN-131107
Code
Text mes-
sage
Cause Rectication
M501
Error in the
non-active
linearization
table
Index markers are not con-
tinuously rising, for example
illogical value pairs
Checklinearizationtable
Delete table/Create new
M502
Error in the
event memory
Hardware error EEPROM Exchanging the electronics
Send instrument for repair
M504
Error on an
device inter-
face
Hardware defect Check connections
Exchanging the electronics
Send instrument for repair
M507
Error in the
instrument
settings
Error during setup
Error when carrying out a
reset
Carry out reset and repeat
setup
9.4 Rectify faults
The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
Therstmeasuresare:
Evaluation of fault messages, for example via the display and
adjustment module
Checking the output signal
Treatment of measurement errors
Further comprehensive diagnostics options are available with a PC
with PACTware and the suitable DTM. In many cases, the reasons can
bedeterminedinthiswayandfaultsrectied.
Connect a multimeter in the suitable measuring range according to
the wiring plan. The following table describes possible errors in the
current signal and helps to remove them:
Error Cause Rectication
4 … 20 mA signal
not stable
Fluctuations of
the measured
variable
Set damping according to the
instrument via the display and
adjustment module or PACTware/
DTM
Reaction when malfunc-
tions occur
Procedure for fault recti-
cation
Check the 4 … 20 mA
signal
Seitenansicht 45