14
4 Mounting
VEGABAR 81 • Modbus and Levelmaster protocol
46293-EN-131107
1 2 3
6
6
6 6
6
5
4
Fig. 5: Position of the lter element - non-Ex, Ex-ia and Ex-d-ia version
1 Single chamber housing plastic, stainless steel (precision casting)
2 Single chamber housing Aluminium
3 Single chamber housing, stainless steel electropolished
4 Double chamber housing plastic
5 Double chamber housing Aluminium
6 Filter element
With the following instruments a blind plug is installed instead of the
lterelement:
Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-
lariesinxconnectedcable
Instruments with absolut pressure
Thelterelementisintegratedintheprocessassembly.Itislocatedin
a rotatable metal ring and has the following function:
Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
Turnthemetalringinsuchawaythatthelterelementpointsdown-
ward after installtion of the instrument. This provides better protection
against buildup.
Instruments in Ex-d ver-