Vega VEGADIF 65 4 … 20 mA Bedienungsanleitung Seite 63

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 88
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 62
63
8Maintenanceandfaultrectication
VEGADIF 65 • 4 … 20 mA
36236-EN-130417
8 Maintenanceandfaultrectication
8.1 Maintenance
Iftheinstrumentisusedproperly,nospecialmaintenanceisrequired
in normal operation.
In some applications, product buildup on the separating diaphragms
caninuencethemeasuringresult.Dependingonthesensorand
application, take precautions to ensure that heavy buildup, and espe-
cially a hardening thereof, is avoided.
8.2 Rectifyfaults
The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
VEGADIF65oersmaximumreliability.Nevertheless,faultscanoc-
curduringoperation.Thesemaybecausedbythefollowing,e.g.:
Sensor
Process
Voltage supply
Signal processing
Therstmeasurestobetakenaretochecktheoutputsignalsaswell
as to evaluate the error messages via the display and adjustment
module.Theprocedureisdescribedbelow.Furthercomprehensive
diagnosticscanbecarriedoutonaPCwiththesoftwarePACTware
andthesuitableDTM.Inmanycases,thecausescanbedetermined
andthefaultsrectiedthisway.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
Thehotlineisavailabletoyou7daysaweekround-the-clock.Since
weoerthisserviceworld-wide,thesupportisonlyavailableinthe
English language. The service is free of charge, only the standard
telephonecostswillbecharged.
Connect a handheld multimeter in the suitable measuring range ac-
cordingtothewiringplan.
Error code Cause Rectication
4 … 20 mA
signal not
stable
Leveluctuations Set the integration time via the
display and adjustment module or
PACTwa re
Maintenance
Reactionwhenmalfunc-
tions occur
Failure reasons
Faultrectication
24 hour service hotline
Check the 4 … 20 mA
signal
Seitenansicht 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 87 88