10 Supplement
10.1 Technical
data
General data
316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
- Process tting thread G1½ A and 1½ NPT: PVDF, Tri-Clamp 2"
and 3", bolting DN 50 and DN 80: adapter of 316L
- Antenna PVDF
- Seal thread FKM (Viton)
- Seal, bolting FKM (Viton)
Materials, non-wetted parts
- Housing Aluminium die-casting powder coated
- Seal between housing and housing
cover
NBR (stainless steel housing), silicone (Alu/plastic
housing)
- Inspection window in housing cover for
PLICSCOM
Polycarbonate (UL-746-C listed)
- Ground terminal 316Ti/316L
Weight depending on process tting and
housing material
0.7 3.4 kg (1.543 7.496 lbs)
Output variable
Output signal 4 20 mA/HART (active)
Cycle time min. 1 s (dependent on the parameter setting)
Signal resolution 1.6 µA
Failure signal current output (adjustable) mA-value unchanged 20 .5 mA, 22 mA, < 3.6 mA
(adjustable)
Max. output current 22 mA
Load < 500 Ω
3)
Damping (63 % of the input variable) 0 999 s, adjustable
Fullled NAMUR recommendations NE 43
HART output values
- 1. HART value (Primary Value) Distance to the le vel
- 2. HART value (Secondary Value) Distance to the le vel - scaled (for example hl, %)
Resolution, digital > 1 mm (0.039 in)
Input variable
Measured variable distance between process tting and product sur-
face
3)
With inductive load ohmic share min. 25 Ω/mH.