6 Set up the sensor with the indicating and adjustment module
VEGAPULS 62 • Modbus and Levelmaster protocol
41362-EN-121011
6 Set up the sensor with the indicating and
adjustment module
6.1 Adjustment volume
The indicating and adjustment module is only used for parameter
adjustment of the sensor, i.e. for adaptation to the measurement task.
The parameter adjustment of the Modbus interface is carried out via
a PC with PACTware. You can nd the procedure in chapter "Set up
sensorandModbusinterfacewithPACTware".
6.2 Insert indicating and adjustment module
The indicating and adjustment module can be inserted into the sen-
sor and removed any time. Four positions displaced by 90° can be
selected. It is not necessary to interrupt the power supply.
Proceed as follows:
1. Unscrew the housing cover
2. Place the indicating and adjustment module in the requested
position onto the electronics and turn to the right until it snaps in
3. Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
Fig. 30: Insertion of the indicating and adjustment module with single chamber
housing
Note:
If you intend to retrot the instrument with an indicating and adjust-
ment module for continuous measured value indication, a higher
cover with an inspection glass is required.