25
4Mounting
VEGADIF 65 • 4 … 20 mA
36236-EN-130417
1. MountVEGADIF65directlytothevessel
2. Connectminussidealwaysabovethemax.level
3. Formeasurementsinproductswithsolidcontentsuchase.g.
dirtyliquids,theinstallationofseparatorsanddrainvalvesis
recommended to enable collection and removal of debris and
sediment.
+
–
max.
min.
+ –
2
3
4
Fig. 23: Measurement setup, level measurement in closed vessel
1 Blocking valve
2 Separator
3 Drain valve
4 VEGADIF 65
1. MountVEGADIF65belowthelowerchemicalseal
2. The ambient temperature should be the same for both capillaries
Information:
Levelmeasurementisonlyensuredbetweentheupperedgeofthe
lowerandtheloweredgeoftheupperchemicalseal.
+
–
min.
max.
1
Fig. 24: Measurement setup, level measurement in closed vessel
1 VEGADIF 65
1. MountVEGADIF65belowthelowermeasurementconnectionso
thattheeectivepressurelinesarealwayslledwithliquid
2. Connectminussidealwaysabovethemax.level
3. The condensate vessel ensures a constant pressure on the
minus side
In closed vessels with
single chemical seal
In closed vessels with
double chemical seal
In closed vessels with
steamlayeringwitheec-
tive pressure line